Agetal AgTIY - RSDetect Equipment Pack BoxAgetal AgTIY - RSDetect Equipment Pack Box
Agetal AgTIY - RSDetect Equipment Pack ContentsAgetal AgTIY - RSDetect Equipment Pack Contents
Agetal AgTIY - RSDetect Equipment Test Kit BoxAgetal AgTIY - RSDetect Equipment Test Kit Box
Agetal AgTIY - RSDetect Equipment Test Kit ContentsAgetal AgTIY - RSDetect Equipment Test Kit Contents

Fast, simple, reliable testing, with no lab required!

Developed in collaboration with AgForce, AgEtal’s AgTIY - RSDetect Test Kits are designed to make the Ratoon Stunting Disease (RSD) testing more accessible for cane growers and reduce the risk of undetected spread in the sugarcane industry.
These on-farm kits enable fast, accurate, and cost-effective detection of Leifsonia xyli subsp. xyli, the bacterium that causes RSD, supports early intervention and effective disease management directly on the farm.

 


 

Why Test for RSD?

RSD is one of the most significant hidden yield-loss diseases in sugarcane; It spreads easily through harvesting equipment and infected planting material, often showing no visible symptoms until a drop in yield is noticed. Early detection is crucial for limiting its spread and maintaining productivity.
AgEtal’s AgTIY - RSDetect Test Kits bring lab-quality diagnostics straight to the farm. Designed specifically for sugarcane growers, these easy-to-use kits provide fast and accurate results without lengthy lab turnaround times.
Each test uses a DNA-based, colour-change method that detects the Leifsonia xyli subsp. xyli pathogen in the cane xylem fluid. Results are visible to the naked eye within an hour — no specialist equipment required!

 


 

What’s Included in the AgTIY - RSDetect Test Kits

*First-time buyers will need to purchase one Equipment Pack. You can then include more Test Kits in the same order.

If the order is over 10 test kits, contact us today. 

Contact

Equipment Pack

Everything you need to begin testing:

  • Micropipette and stand
  • Micropipette tips (box of 96 tips)
  • Pipetting practice tubes and tips
  • Heat block
  • Tube racks x3 (sample rack, reagent rack and test rack)
  • Written instructions
  • Box of ethanol wipes
  • Sharpie pen


Option: Glove Size

Test Kit

The AgTIY - RSDetect Test Kits are Ideal for routine monitoring or multi-site testing. The AgTIY - RSDetect Test Kits contain consumables for one batch of testing (up to six samples):

  • Sample collection tubes
  • Reaction tubes and detection reagents
  • Test reagents (colour-change indicator dyes included)
  • Optional extras (gloves, medical wipes and pipette tips)

 

Why Choose AgEtal’s AgTIY - RSDetect Test Kits

  • Prevent yield loss
    Early detection lets you remove or manage infected cane before yield is impacted.
  • Improve on-farm biosecurity
    Avoid the spread of RSD via contaminated planting material or machinery.
  • Fast results
    No need to wait days for lab results.
  • Supports industry best practice
    Enables testing of mother plots, planting material & suspect blocks.
  • AgForce-endorsed
    Marketed in partnership with AgForce Queensland to build diagnostic capacity among growers & contractors.

 


 

Everything you need to know before testing your cane

 

Handling samples on testing day

Sample tubes have the potential to cause cross-contamination between samples. The outside of the tube may have been contaminated with the sample containing bacteria during collection, which can then contaminate gloved hands and/or equipment.

It is recommended to handle samples in the following manner:

  • Use a clean and dry sample tube rack.
  • Put on a fresh pair of gloves.
  • Check that the sample tube lid is closed firmly.
  • Then mix the sample by swirling the tube strongly.
  • Wipe the outside of the sample tube using an ethanol wipe.
  • Guide the sample down into the bottom of the tube by gently tapping the tube on the bench.
  • Change gloves before handling the next sample tube.
  • Leave samples to sit for 15 min before testing to allow the plant material to settle to the bottom.

RSD Test Kits hygiene & safety

It is important to follow some general hygiene and safety guidelines for your protection and for the protection of others. However, it is also important to avoid actions that might compromise the test results through cross-contamination from sample to sample or from the environment to the samples.

Follow the steps in the same order to ensure the best protection of the testing integrity.

  • If you have long hair, tie it back to keep it away from your face and samples and equipment.
  • Wear appropriate protective clothing such as a clean long-sleeved shirt or a clean dedicated coat.
  • We also encourage the use of close-fitting safety glasses to protect your eyes.
  • Then put on a pair of gloves.
  • Once you have put on your gloves, you should avoid touching unnecessary items. For example, don't touch your skin, hair, the chair, the ground, or door handles with your gloves. If anything drops on the ground, leave it there until the end of the testing.
  • Clean gloves with an ethanol wipe and dispose of the ethanol wipe in a plastic bag. You could use an empty bread bag for this purpose.
  • Wipe the bench area with an ethanol wipe before the testing.
  • Wipe down the tube racks, the outside surface of the tip box, and the micropipette with ethanol wipes.
  • Perform the testing following the test instructions. Refer to the accompanying video and written instructions for the testing.
  • All used or contaminated tips should be disposed of into a disposal bottle with a lid that can be closed before disposing of the bottle into general rubbish bins. You could use an empty milk bottle for this purpose.
  • At the end of the testing, dispose of the reagent tubes into the disposal bottle and close the lid of the bottle.
  • You can also dispose of the sample tubes into the same bottle once you have your results and no longer need the samples.
  • After testing is completed, wipe down the equipment and bench area again with ethanol wipes before putting the equipment away.
  • Dispose of your gloves in the plastic bag. Seal the bag either by tying a knot in the bag or using a rubber band before disposing into general rubbish bins.

Chemical Safety

There are no known hazards identified with the chemicals used in this kit. However, individuals can have varying sensitivities to different reagents, and while only very small quantities are supplied, care should always be taken to avoid exposure of skin and eyes to these chemicals.

When storing the kit reagents, consider the location and manner of storage to ensure they are kept out of reach of children or other individuals who don’t know how to handle them. If contact is made by kit reagents with skin or eyes, flush thoroughly with water.

Micropipette handling

Micropipettes (with a tip on the end) can measure very small volumes of liquid accurately. The micropipette supplied measures 20 microlitres of liquid = 1/50th of a mL.

A box of very clean (sterile) tips has been supplied that matches the micropipette. Always keep the lid of the box closed to keep the tips clean.

It is important to take care of the micropipette to ensure that it continues to measure accurately and to avoid cross-contamination. Micropipettes should always be in an upright position to ensure that liquid does not accidentally enter the micropipette shaft.

Wear gloves and clean the surface of the micropipette with an ethanol wipe at the beginning and at the end of your work. You should also clean with an ethanol wipe if you suspect contamination of the surface has occurred at any stage.

Hold the micropipette in a vertical (upright) position during use, in either hand, with the black tip ejector facing your thumb.

Tubes and tips have been separately supplied for the purposes of practising the use of the micropipette. The spare tips are in the packaging with the micropipette, and the tubes are in a plastic bag in the box with the equipment.Tubes and tips have been separately supplied for the purposes of practising the use of the micropipette. The spare tips are in the packaging with the micropipette, and the tubes are in a plastic bag in the box with the equipment.

Practice pressing down slowly on the white button on the top of the micropipette to the first stop, holding for 2 seconds, and then releasing the button upwards slowly.

Then practice pushing on the white button slowly again to the first stop, and then pushing slowly further to the end, the second stop, hold for 2 seconds, and then release the button upwards slowly.

Next, use the micropipette to measure 20?L of liquid from a sample tube into a reagent tube.

  • Place the sample tube into the sample rack. Place the reagent tube into the reagent rack.
  • Remove the lid of a disposable bottle. You could use an empty milk bottle for this purpose.
  • Loosen the lid of the sample tube.
  • Hold the micropipette in one hand.
  • Open the lid of the tip box with the other hand. Do not touch the tips.
  • Gently insert the bottom end of the micropipette into the top of a tip. Tap gently to check the tip stays on.
  • Lift the pipette away with the tip on the end and close the tip box.
  • Using your free hand, place the sample tube lid upside down on the bench. Now, pick up the open sample tube without touching the top and inner edge of the sample tube.
  • Press the white button on the top of the micropipette down to the first stop and hold.
  • While holding at the first stop, place the tip into the top layer of the liquid in the sample tube.
  • Slowly release the white button upwards to draw up the liquid.
  • Lift the tip out of the tube.
  • Place the tube back into the rack and set the lid on top.
  • Pick up the reagent tube and carefully open it using your thumb to press upwards on the lid. Avoid touching any part of the lid that goes inside the tube.
  • Place the tip inside the reagent tube and just into the top of the reagent liquid.
  • Press the white button slowly to the first stop to eject the measured liquid.
  • Mix the contents by slowly releasing and slowly pressing down the white button 4x.
  • At the last downward button press to the first stop, hold, and then lift the tip just clear of the liquid.
  • Touch the tip to the inner wall of the tube just above the liquid and press the button further through to the second stop and hold to eject all the remaining liquid onto the tube wall.
  • Move the tip off the wall of the tube whilst holding the button at the second stop.
  • Slowly release the white button and lift the tip clear of the tube.
  • Close the tube lid without touching any internal parts and place the tube back into the reagent rack.
  • Discard the tip into the disposal bottle by holding the tip just inside the neck of the disposal bottle and pressing the black tip ejector. The tip should drop off into the bottle.
  • Place the micropipette back on the stand.

Always use a new tip between samples and for each reagent.

Be aware of where the tip touches. If you bump a rack, the bench, or the outside of a tube, then you must discard that tip.

If a used tip falls onto the bench, leave it there until the end of the testing.

Do not leave tips on the end of the micropipette when on the stand, as they will not stay clean, or they might get used twice, creating contamination across samples and/or tests.

Testing Instructions


The order of pipetting should generally be from clean samples that have no or little bacteria to dirty samples that contain lots of bacteria. Hence, the negative control tubes (shown by the green dots) should be the first handled at each step, and the positive control tubes (shown by the red dots) should be last at each step.

  • Prepare yourself and the testing area as outlined in the Hygiene and Safety section.
  • Each test strip can test up to 6 sugarcane xylem samples.
  • Prepare sample tubes as per the Handling samples section. Mix each sugarcane xylem sample tube and then allow them to stand for 15 min in the middle positions of the front row in the Sample rack (see the Sample rack diagram below). The positions at either end (red and green) is where the control sample tubes will be sitting later.
  • Turn on the heat block to 67°C.
    • The heat block will have been pre-set before delivery. If not, follow the instructions in the booklet provided with the heat block.
  • Complete the sample order table on the results sheet. Spare results sheets can be downloaded here
    Results Sheet
  • Remove all reagent bags from the fridge and the freezer.
    • 1x bag of (two) control tubes (from fridge)
    • 1x bag of (eight) Reagent 1 tubes (from fridge)
    • 1x bag of (eight) Reagent 2 tubes (from fridge)
    • 1x bag of (eight) Test reaction tubes (from freezer)
  • Open the Control sample tubes and place them into the Sample rack.
  • Check that the control and sample vials are in the same order as the table on the results sheet.
  • Open the bags for Reagents 1 and 2 and place them into the Reagent rack (see the suggested positioning in the Reagent rack diagram below).
  • Remove one packet containing 1 strip of 8 tubes of freeze-dried reaction mix from the freezer. Allow the pack to come up to room temperature while processing samples before opening.
  • Samples should be processed through the buffers fairly promptly, with the time in buffer 1 no longer than 5 min. They can be processed either individually or as a group through the buffers.
  • Working from right to left:
    • Add 20?L of the sample into Buffer 1
    • Pipette gently up and down 4x to mix.
    • Close the lid and mix by gently flicking the bottom of the tube with a finger.
  • Again, working from right to left:
    • Add 20?L of the “sample/Buffer 1 mix” into Buffer 2
    • Pipette gently up and down 4x to mix.
    • Close the lid and mix by gently flicking the bottom of the tube with a finger.
  • Check that the temperature of the heat block has reached 67°C.
  • Open the packet of freeze-dried reaction mix tubes and place the tubes into the Test rack (see the Test rack diagram below). Carefully close any lids that may have popped open without touching the internal parts.
  • Again, working from right to left:
    • Add 20?L of the “sample/Buffer 2 mix” to the reaction tubes.
    • Flick gently to mix.
  • Leave tubes to sit for 3 min, flicking gently every min
  • Check that all the lids are firmly closed.
  • Up-end the tube strip sharply 5x to mix and also to catch any pink reagent that might be sitting further up the tubes.
  • Ensure all liquid has returned to the bottom of the tubes by flicking the strip downwards.
  • Place the tubes into the heat block at 67°C.
  • Start a timer for 40 minutes.
  • At the end of the 40-minute incubation period, open the lid of the heat block carefully and check that all tube lids are firmly closed.
  • It is very important that these incubated tubes are not opened, as the levels of DNA will have increased substantially and can be a source of contamination to the surrounding area and equipment, affecting future test results. If this issue occurs, follow the decontamination protocol below.

  • Remove the strip of tubes from the heat block, and place it on a white background to view the colour results.
  • Check that the negative control tube has remained a pink/red colour. If the negative control has changed to a yellow or orange colour, then the results are compromised, possibly by cross-contamination. Either way, record the result on the results sheet. In the case of compromised results, the best action is to change gloves and thoroughly clean the area, the equipment and the outside of the sample tubes before repeating the entire set of controls and samples. If a second set of compromised results occurs, contact AgEtal for advice.
  • Check that the positive control tube has changed to a yellow colour. If the positive control has not changed to yellow, then the results are also compromised, as this result suggests that the reagents have not worked. Either way, record the result on the results sheet.
  • Then record the results for each sample. If the controls have worked correctly, then any sample that is orange should be treated as a possible weak positive result and can be sent to AgEtal for confirmation.

Decontamination protocol

Wipe all benches and racks with a dilute bleach solution and then wipe again with water containing a few drops of dishwashing detergent (detergent water). Turn the power off to the heat block and let it cool down.

Once cool, wipe the heat block with ethanol wipes, then wipe again using a cloth dampened with detergent water and then wipe a third time with ethanol wipes to reduce the detergent load on the metal surfaces of the equipment.

Use cotton tips dampened with detergent water to clean gently inside each well of the heat block, and then clean wells with cotton tips dampened with water only to reduce the detergent load on the metal surfaces.

 

 

Contact Us!

Have a question about our on-farm RSD Test Kits? Reach out to us today!

Contact Us

 

 

 


Would you prefer to Print or Download this form?


If you would prefer not to complete this form online, you can download or print the form.

Simply complete the form on paper, attach to your sample and post or deliver to AgEtal.

If you require testing to full ISTA methods (i.e. 4 reps instead of 2) or for OIC/BIC please use the ISTA form.

ISTA International Form

Download Domestic Analysis Form

New Customer Form


Accurate results
in a timely manner.


 

Request for Analysis.

Simply book online for all your seed and grain testing then drop off or post the sample to our lab.

Please be sure about the tests you require before submitting your Request for Analysis. *See our Cancellation Policy
Test results have a fast turnaround and are available online.

Request for your samples to be analysed in three easy steps.


Complete the
form below.
Sign & attach the receipt
to your sample.
Post or deliver the
sample to AgEtal.


Attach the form to your sample. Post or deliver the sample to AgEtal. For fast turnaround, please deliver samples before 3:30pm.

Thank you for requesting an analysis

We have received your request for analysis, don't forget to print and sign the receipt below to attach to your samples. Send or deliver your samples to our laboratory. If you have any questions, please contact us.

We will be in contact with you when your results are ready.

Print Receipt

Attach the form to your sample. Post or deliver the sample to AgEtal.

AgEtal Request Analysis

Customer Details
Accounts Payable Details
Analysis Details 1
Lab. No. (Office use only)
Analysis Details 2
Lab. No. (Office use only)
Analysis Details 3
Lab. No. (Office use only)
Analysis Details 4
Lab. No. (Office use only)
Analysis Details 5
Lab. No. (Office use only)
Analysis Details 6
Lab. No. (Office use only)
Analysis Details 7
Lab. No. (Office use only)
Analysis Details 8
Lab. No. (Office use only)
Accredited Sampler & Other

* Abbreviations: Purity = P  Germination = G  Vigour = V  Tetrazolium = Tz  |  Weed Search = WS


A booking receipt will generate when submiting this form. (a confirmation will sent to the nominated Customer email)

Test results have a fast turnaround and are available online.

Please remember to complete the final steps when you submit this form.

Sign & attach the receipt
to your sample.

(Booking Receipt)
 Post or deliver the
sample to AgEtal.

Export Information (if applicable)
Analysis Submit
Issued by Luanne Cunliffe Director Webform WAM03 01/09/21

Contact AgEtal Agricultural Testing Laboratory

Get in contact with our laboratory to book an analysis or learn more about our testing.


Phone 07 4633 3223

[email protected]

Courier 9/24 Carroll St Toowoomba QLD 4350

Auspost PO Box 7135 Toowoomba South 4350



Opening Hours

Monday to Friday
8:30am to 5:00pm

Saturday & Sunday
Closed to the public


Counting is done every day – after hours drop off box available.


Follow us on Social Media

 

 

Send a message

Need some more information about the laboratory services we provide, have a technical question, feedback or want to discuss an idea with us?

Select a form and forward your message. 

General Enquiry

AgEtal Contact

Send a message

Technical Questions

Agetal - Technical Questions

Technical Questions

Customer Feedback

Agetal - Customer Feedback

Customer Feedback

Subscribe to our mailing list

* indicates required